Hp83k down

    Faint, Only after Agilent did the calibration several days. Consider with shipping, they skip hp83k test ,direct goto write SPD.
Error message:
    
No calibration data has been download to the tester hardware
------------------------------------------------------------
Servere Environment Error or program bug.
Either the MCD has errors or the calibration data are correput or out of date.
------------------------------------------------------------
 
display:
No device specified in this operator's workorders are accessible on this tester!
If the devices should be mounted please make sure that the server is up and running.
To check which devices are accessible please start the software in engineer or superior mode.

    My Abg  went to  check the program through network, and reply my mail tell me what to do.

What you should do is re-do the Cal ,they was file missing in calibration dir you can check following dir :-

  $   cd /opt/hp83000/f330/fw/data/calibration

  $   ll std__

        std__ not found

You should have this std__” file to run the tester .Also the “pdps “file is corrupted .

Please following this step to re-do the cal .

Step 1 :-

  1. open dtterm and type command as below “don’t logging as root “:-

$  cd/opt/hp83000/f330/pws/bin

$ ./hp83000 &

      Wait until system going into debug mode “Engineering Mode “

Step 2 :-

  1. Please do the CAL as normal and save the calibration file .
  2. Start the tester as normal

    I hope it will helpful for run the smarttest,yesterday I tried to do the calibration when the operator tell me that hp83k had some problem, but it appeared 2 two channel is bad! but next cal, the error was gone. 
    Two days ago, the water coupler for  DC DC Converter Card was bad, and water flowed all the floor......
    
   